Chapter 3
61
Operating Overview
Calibrating the Channel Power Measurement – CDMA Analyzer
Calibrating the Channel Power Measurement
– CDMA Analyzer
– CDMA Analyzer
To calibrate a channel power measurement using the internal CDMA
generator:
generator:
Step 4. Remove power from the RF IN/OUT (or ANT IN) port.
Step 5. Go to the CDMA ANALYZER screen and choose the
Chan Pwr
measurement.
Step 6. Set the
Tune Freq
field (or
RF Channel
field) to the center frequency of
your CDMA signal.
Step 7. Select the
Chn Pwr Cal
field and wait until the calibration is complete.