Chapter 3
69
Operating Overview
Setting an RF Generator/Analyzer Offset
Setting an RF Generator/Analyzer Offset
You can set a fixed frequency offset between the RF generator and the
RF analyzer. This feature is convenient for testing radios with a fixed
transmit/receive frequency offset.
RF analyzer. This feature is convenient for testing radios with a fixed
transmit/receive frequency offset.
To set an RF Offset
Step 1. Go to the INSTRUMENT CONFIGURE screen.
Step 2. Set the
RF Display
field to
Freq
.
Step 3. Set the
RF Offset
field to
On
.
Step 4. Select the
(Gen)-(Anl)
field and enter the frequency offset value.
Step 5. Go to the RF GENERATOR screen.
Step 6. Select the
RF Gen Freq
field, and rotate the cursor control knob to vary
the RF generator’s frequency.
Step 7. Go to the RF ANALYZER screen. Notice that the
Tune Freq
value
changes to maintain the offset between the generator and the analyzer.
(You can use the
(You can use the
Prev
key to switch between the generator and analyzer
screens.)