Chapter 3
69
Operating Overview
Setting an RF Generator/Analyzer Offset
Setting an RF Generator/Analyzer Offset 
You can set a fixed frequency offset between the RF generator and the 
RF analyzer. This feature is convenient for testing radios with a fixed 
transmit/receive frequency offset. 
To set an RF Offset 
Step 1. Go to the INSTRUMENT CONFIGURE screen. 
Step 2. Set the 
RF Display
 field to 
Freq
.
Step 3. Set the 
RF Offset
 field to 
On
Step 4. Select the 
(Gen)-(Anl)
 field and enter the frequency offset value. 
Step 5. Go to the RF GENERATOR screen. 
Step 6. Select the 
RF Gen Freq
 field, and rotate the cursor control knob to vary 
the RF generator’s frequency. 
Step 7. Go to the RF ANALYZER screen. Notice that the 
Tune Freq
 value 
changes to maintain the offset between the generator and the analyzer. 
(You can use the 
Prev
 key to switch between the generator and analyzer 
screens.)