Agilent Technologies 8935 series e6380a Manuale Utente
Chapter 4
89
Overview of the Test Set’s Built-In Tools
Using the RF Tools Program
Swept Insertion Loss
Measurement Description
This test measures the loss of a cable or device by sweeping it over the
specified frequency range. This test uses the Spectrum Analyzer to
make measurements.
specified frequency range. This test uses the Spectrum Analyzer to
make measurements.
Making the Measurement
1. Select
Swept Insertion Loss
from the Select Test (Main) Menu. A
Swept Insertion Loss Information menu will appear.
Figure 4-10
Selecting the Swept Insertion Loss Measurement
2. Position the cursor at
Start frequency in MHz
and select it. Enter
a value for the start of the frequency sweep from 0.4 MHz to
1000 MHz or 1700 MHz to 2000 MHz.
1000 MHz or 1700 MHz to 2000 MHz.
3. Position the cursor at
Stop frequency in MHz
and select it. Enter a
value from 0.4 MHz to 1000 MHz or 1700 MHz to 2000 MHz. This
value should be larger than the start frequency.
value should be larger than the start frequency.
Swept Insertion Loss
Select Test Menu