Agilent Technologies E6380-90073 Manuale Utente
110
Chapter 1
GPIB Command Dictionary
CANalyzer subsystem
:POWer:GAIN:MODE <string>
:POWer:GAIN:MODE?
:POWer:GAIN:MODE?
These commands set/query the mode of the
Pwr Gain
field, allowing
you to choose between automatic gain (Auto) or fixed gain (Hold).This
command is valid only when average power is selected. For channel
power, EVM and rho measurements, use the CAN:PATH:GAIN:MODE
command.
command is valid only when average power is selected. For channel
power, EVM and rho measurements, use the CAN:PATH:GAIN:MODE
command.
Syntax
CAN:POW:GAIN:MODE ‘Auto’
‘Hold’
CAN:POW:GAIN:MODE?
Screen/field equivalent
CAN:POW:GAIN:MODE sets the
Auto/Hold
subfield of the
Pwr Gain
field of the CDMA ANALYZER screen, when
Avg Pwr
is displayed.
:POWer:SAMPle:TIME <real number>
:POWer:SAMPle:TIME?
:POWer:SAMPle:TIME?
These commands set/query the power interval. This interval is the
amount of time that average power or channel power is measured.This
command utilizes the “Real Number Setting Syntax” on page 299, but
does not use the :STATe command.
amount of time that average power or channel power is measured.This
command utilizes the “Real Number Setting Syntax” on page 299, but
does not use the :STATe command.
Syntax
CAN:POW:SAMP:TIME <real number>
!values from 0.00025 to .02666 s
CAN:POW:SAMP:TIME?
Screen/field equivalent
CAN:POW:SAMP:TIME sets the
Pwr Intvl
field of the CDMA
ANALYZER screen.