GPIB Command Dictionary
CDANalyzer subsystem
:CPOWer:CALibrate
This command initiates the channel power calibration. When this 
command is received, an internally generated calibration signal is 
measured using the average power technique. Channel power is also 
measured and a correction factor is generated. This correction factor is 
applied to subsequent channel power measurements. Calibration 
should be performed whenever a new set of measurements is made and 
whenever the frequency of the measured signal is changed.
Syntax
CDAN:CPOW:CAL
Screen/field equivalent  
CDAN:CPOW:CAL controls 
Chn Pwr Cal
 measurement field on the 
CD 
Setup 
menu of the CODE DOM screen, when the CDMA standard is 
set to “IS-95 Only”. The field displayed on the 
CD Setup 
menu when 
the 
Measurement 
field on the 
Main 
menu is set to 
Power 
or 
Fast 
Power
, and the 
CD pwr unit 
field on the 
CD Setup 
menu is set to 
Abs
.
:EVENt:QUALifier <string>
:EVENt:QUALifier?
These commands set/query the qualifying event for the trigger. The 
qualifying event is the event that must occur before a trigger is 
accepted. The trigger event is set by the :EVEN:TRIG command.
Syntax
CDAN:EVEN:QUAL ‘None’
‘27 ms’
 ‘20 
ms’
 ‘80 
ms’
 
‘2 s’
 ‘Ampl 
Lo’
 ‘Ampl 
Mid’
‘Ampl Hi
‘External’
Screen/field equivalent
CDAN:EVEN:QUAL controls the 
Qual Event
 field of the CODE DOM 
screen. This field is displayed on the 
Trigger
 menu, when the CDMA 
standard is set to “IS-95 Only”.