GPIB Command Dictionary
CDANalyzer subsystem
:CPOWer:CALibrate
This command initiates the channel power calibration. When this
command is received, an internally generated calibration signal is
measured using the average power technique. Channel power is also
measured and a correction factor is generated. This correction factor is
applied to subsequent channel power measurements. Calibration
should be performed whenever a new set of measurements is made and
whenever the frequency of the measured signal is changed.
command is received, an internally generated calibration signal is
measured using the average power technique. Channel power is also
measured and a correction factor is generated. This correction factor is
applied to subsequent channel power measurements. Calibration
should be performed whenever a new set of measurements is made and
whenever the frequency of the measured signal is changed.
Syntax
CDAN:CPOW:CAL
Screen/field equivalent
CDAN:CPOW:CAL controls
Chn Pwr Cal
measurement field on the
CD
Setup
menu of the CODE DOM screen, when the CDMA standard is
set to “IS-95 Only”. The field displayed on the
CD Setup
menu when
the
Measurement
field on the
Main
menu is set to
Power
or
Fast
Power
, and the
CD pwr unit
field on the
CD Setup
menu is set to
Abs
.
:EVENt:QUALifier <string>
:EVENt:QUALifier?
:EVENt:QUALifier?
These commands set/query the qualifying event for the trigger. The
qualifying event is the event that must occur before a trigger is
accepted. The trigger event is set by the :EVEN:TRIG command.
qualifying event is the event that must occur before a trigger is
accepted. The trigger event is set by the :EVEN:TRIG command.
Syntax
CDAN:EVEN:QUAL ‘None’
‘27 ms’
‘20
ms’
‘80
ms’
‘2 s’
‘Ampl
Lo’
‘Ampl
Mid’
‘Ampl Hi
‘External’
Screen/field equivalent
CDAN:EVEN:QUAL controls the
Qual Event
field of the CODE DOM
screen. This field is displayed on the
Trigger
menu, when the CDMA
standard is set to “IS-95 Only”.