Agilent Technologies E6380-90073 Manuale Utente
Chapter 1
137
GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:MEASurement:MODE <string>
:IS2000:MEASurement:MODE?
:IS2000:MEASurement:MODE?
IS-2000 only. These commands set/query the measurement state for the
code domain analyzer. This mode can be overridden with the
TRIG:MODE:RETR REP and TRIG:MODE:RETR SING trigger
commands. When the trigger commands are used, the state of the field
on the display are not updated. For example, the screen may show
code domain analyzer. This mode can be overridden with the
TRIG:MODE:RETR REP and TRIG:MODE:RETR SING trigger
commands. When the trigger commands are used, the state of the field
on the display are not updated. For example, the screen may show
Single
selected when TRIG:MODE:RETR REP is used.
Syntax
CDAN:IS2000:MEAS:MODE ‘Single’
‘Cont’
CDAN:IS2000:MEAS:MODE?
Screen/field equivalent
CDAN:IS2000:MEAS:MODE controls the
Single/Cont
subfield of the
Measurement
field of the
Main
image of the CODE DOMAIN screen
when the CDMA standard is set to “IS-2000”.
:IS2000:MEASurement:ORDer <string>
:IS2000:MEASurement:ORDer?
:IS2000:MEASurement:ORDer?
IS-2000 only. These commands set/query the channel order of
measurement currently shown on the CODE DOMAIN screen. The
channel order can follow the Hadamard ordering in which all 128
channels are ordered consecutively from 0 to 127, or they can follow the
Bit Reverse ordering.
measurement currently shown on the CODE DOMAIN screen. The
channel order can follow the Hadamard ordering in which all 128
channels are ordered consecutively from 0 to 127, or they can follow the
Bit Reverse ordering.
Syntax
CDAN:IS2000:MEAS:ORD ‘Bit Reverse’
‘Hadamard’
CDAN:IS2000:MEAS:ORD?
Screen/field equivalent
CDAN:IS2000:MEAS:ORD controls the
Meas Order
field of the
Aux
image on the CODE DOMAIN screen when the CDMA standard is set
to “IS-2000”.
to “IS-2000”.