Chapter 1
141
GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:POWer:FAST:NAVG <integer>
:IS2000:POWer:FAST:NAVG?
IS-2000 only. These commands set/query the number of averages made 
during fast power measurements.
Syntax
CDAN:IS2000:POW:FAST:NAVG <integer>
CDAN:IS2000:POW:FAST:NAVG?
Screen/field equivalent
CDAN:IS2000:POW:FAST:NAVG controls the 
Num Avgs
 field of the 
FP 
Setup 
menu on the CODE DOMAIN screen when the CDMA standard 
is set to “IS-2000”.
:IS2000:POWer:UNIT <string>
:IS2000:POWer:UNIT?
IS-2000 only. These commands set/query whether power measurements 
are absolute power measurements or relative power measurements.
• ‘Abs’ - Absolute code domain power displays the power in each of the 
128 Walsh channels, relative to the total power inside a 1.23 MHz 
bandwidth centered at the tune frequency. 
• ‘Rel’ - Relative code domain power displays the power in each of the 
128 Walsh channels, relative to the power of the Pilot signal. (Pilot 
power is approximately two-thirds of the total power.)
Syntax
CDAN:IS2000:POW:UNIT ‘Abs’
‘Rel’
CDAN:IS2000:POW:UNIT?
Screen/field equivalent
CDAN:IS2000:POW:UNIT controls the 
CD pwr unit
 field of the 
Reference 
image on the CODE DOMAIN screen when the CDMA 
standard is set to “IS-2000”.