CGENerator subsystem
:EBNO:LEVel <real number>
:EBNO:LEVel?
These commands set/query the level of the generator when it is in 
Eb/No mode. This command utilizes the “Real Number Setting Syntax” 
on page  299,
 but does not use the :STATe command.
Syntax
CGEN:EBNO:LEV <real number> !values from 
5 to +25 dB
CGEN:EBNO:LEV?
Screen/field equivalent
:EBNO:LEV controls the generator’s level when the 
Gen Mode
 field on 
the CDMA GENERATOR screen is set to 
Eb/No
.
:EQFilter <string>
:EQFilter?
These commands set/query the setting of the equalization filter used to 
create forward channel signals. This command is used when you want 
to override the default setting (EQ In). The filter is bypassed for reverse 
link signals and cannot be overridden. Therefore, this command is valid 
only when the CGEN:DIR ‘Fwd’ command is used.
Syntax
CGEN:EQF ‘EQ In’ !switches the Equalization filter in
 
‘Out’ 
!bypasses the Equalization filter
CGEN:EQF?
Screen/field equivalent
CGEN:EQF controls the 
EQ In/Out
 subfield of the 
Gen Dir
 field on the 
CDMA GENERATOR screen. This field is displayed only when 
Fwd
 is 
selected in the 
Fwd/Rev