Chapter 1
 
GPIB Command Dictionary
DECoder subsystem
:NAMPs|NTACs:STANdard <string>
:NAMPs|NTACs:STANdard?
These commands set/query the expected frame structure and channel 
range of the decoder’s incoming signal. Trying to run a test with the 
wrong standard selected will result in incorrectly decoded data, or will 
result in a displayed error message.
Syntax
DEC:NAMP:STAN ‘NAMP’
 ‘NTAC’
DEC:NAMP:STAN?
Screen/field equivalent
DEC:NAMP:STAN controls the 
Standard
 field on the SIGNALING 
DECODER screen when the 
Mode
 field is set to 
NAMP-NTAC
.
:NAMPs|NTACs:TRIGger:PATTern <string>
:NAMPs|NTACs:TRIGger:PATTern?
These commands set/query the bit pattern to filter displayed 
information. The decoder only displays the received data when this 
binary pattern is encountered immediately after triggering. This is 
helpful when you only want to display messages containing very 
specific information.
The trigger pattern is entered as a sequence of ones, zeros, and dots. A 
dot will cause the decoder to trigger for either a one or a zero in that bit 
position in the received data stream.
This function is not available for decoding NAMPS-NTACS RVC 
information.
Syntax
DEC:NAMP:TRIG:PATT <string>
DEC:NAMP:TRIG:PATT?
Screen/field equivalent
DEC:NAMP:TRIG:PATT controls the 
Trigger Pattern (bin
) field on 
the SIGNALING DECODER screen when the 
Mode
 field is set to 
NAMP-NTAC
, and the 
Channel
 field is set to 
Cntl
.