Agilent Technologies E6380-90073 Manuale Utente
Chapter 1
209
GPIB Command Dictionary
I/Q Plot Subsystem
:TRIGger:DARM
This command disarms the IQ Plot measurement when it has been put
into single triggering mode using the TRIG:MODE:RETR SING. See
“TRIGger subsystem” on page 346 for a more complete discussion of
GPIB triggering.
into single triggering mode using the TRIG:MODE:RETR SING. See
“TRIGger subsystem” on page 346 for a more complete discussion of
GPIB triggering.
Syntax
IQPL:TRIG:DARM !disarms the IQ Plot measurements.
Screen/field equivalent
IQPL:TRIG:DARM controls the
Disarm
selection of the
Analyzer
field
of the
Main
image of the IQ PLOT screen.
:TRIGger:DELay <real number>
:TRIGger:DELay?
:TRIGger:DELay?
These commands set/query the current field contents of the
Trig Delay
field on the IQ Plot screen.
Syntax
IQPL:TRIG:DEL <real number from 20
µ
s to 10 s>
IQPL:TRIG:DEL?
Screen/field equivalent
IQPL:TRIG:DEL sets the
Trig Delay
field of the
Trigger
image when
the
Trig Event
field is set to
Delay
on the IQ PLOT screen. The query
returns the current value.
:TRIGger[:EVENt] <string>
:TRIGger[:EVENt]?
:TRIGger[:EVENt]?
These commands set/query the trigger interval in the IQ plot screen.
Syntax
IQPL:TRIG[:EVEN] ‘27 ms’
‘20 ms’
‘80 ms’
‘2 s’
‘Delay’
‘Immed’
IQPL:TRIG[:EVEN]?
Screen/field equivalent
IQPL:TRIG[:EVEN] sets the
Trig Event
field of the
Trigger
image
on the IQ PLOT screen.