Chapter 1
 
GPIB Command Dictionary
MEASure subsystem
:AFRequency:DCVolts <meas cmnd>
:AFRequency:DCVolts?
These commands set/query the DC voltmeter. To use this measurement, 
you must have the DC level measurement selected (see 
MEAS:AFR:SEL on page  216), and have selected one of the following 
audio sources: SSB Demod, Audio In, Ext Mod, or Audio Out (see 
AFAN:INP on page  59).
Syntax
MEAS:AFR:DCV <meas cmd> 
MEAS:AFR:DCV? !returns a real value
Example
MEAS:AFR:DCV mV !sets the measurement unit to mV
MEAS:AFR:DCV? 
!returns the real value in millivolts
Screen/field equivalent
MEAS:AFR:DCV? reads the 
DC Level
 (V) measurement field on the AF 
ANALYZER screen when the 
AF Anl In
 field is set to 
Audio In
Audio 
Out
Ext Mod
, or 
SSB Demod
. This measurement is also displayed on the 
RF GENERATOR and RF ANALYZER screens.
:AFRequency:DISTortion <meas cmnd>
:AFRequency:DISTortion?
These commands set/query the distortion measurement. This 
measurement is selected by using the MEAS:AFR:SEL ‘Distn’ 
command (see page  216), then you can read the measurement results 
using the :AFR:DIST query command.
Syntax
MEAS:AFR:DIST <meas cmd> 
MEAS:AFR:DIST? !returns a real value
Example
MEAS:AFR:SEL ‘DISTN’ ! selects the distortion measurement
MEAS:AFR:DIST? 
!returns the measured value
Screen/field equivalent
MEAS:AFR:DIST? reads the 
Distn
 measurement field on the AF 
ANALYZER screen. This measurement is also displayed on the RF 
GENERATOR and RF ANALYZER screens.