Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
Chapter 1
247
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:ABSolute:POWer:NOISe:ACTive?
IS-2000 only. This query returns active channel information for the 
Code Domain trace when making a Power and Noise measurement. The 
return entity consists of an integer representing the number of active 
channels, followed by a four value record for each channel. The four 
values for each record are channel number, spread factor for this 
channel, Power, and Noise for this channel.
Syntax
MEAS:CDAN:IS2000:TRAC:ABS:POW:NOIS:ACT?
Screen/field equivalent
There are no fields on the CODE DOMAIN screen equivalent to the 
return values for these queries. These values are only accessible when 
the CDMA standard is set to “IS-2000”, the measurement selected is 
Power and Noise
, and 
CD pwr unit
 is set to 
Abs
 on the Reference 
image of the CODE DOMAIN screen.