GPIB Command Dictionary
MEASure subsystem
:OSCilloscope:MARKer:LEVel:VOLTs <meas cmnd>
:OSCilloscope:MARKer:LEVel:VOLTs?
These commands set/query the voltage measurement at the 
oscilloscope’s marker position. For this measurement to be valid, you 
must have Audio In, Audio Out, Ext Mod, or SSB Demod chosen as the 
audio input (see AFAN:INP on page  59).
This measurement utilizes the Number Measurement Syntax, but does 
not use the :METer command.
Syntax
MEAS:OSC:MARK:LEV:VOLT <meas cmd> 
MEAS:OSC:MARK:LEV:VOLT?
Example
MEAS:OSC:MARK:LEV:VOLT? 
Screen/field equivalent
MEAS:OSC:MARK:LEV:VOLT? reads the 
Lvl
 measurement field on 
the SCOPE screen. The 
AF Anl In
 field on the AF ANALYZER screen 
must be set to 
Audio In
Audio Out
, or 
Ext Mod
SSB Demod
 to measure 
voltage in this field.
:OSCilloscope:MARKer:TIME <meas cmnd>
:OSCilloscope:MARKer:TIME?
These commands set/query the time elapsed from the trigger event to 
the marker location. 
This measurement utilizes the Number Measurement Syntax, but does 
not use the :METer command.
Syntax
MEAS:OSC:MARK:TIME <meas cmd> 
MEAS:OSC:MARK:TIME? !returns a real number
Example
MEAS:OSC:MARK:TIME? !displays the measurement 
Screen/field equivalent
MEAS:OSC:MARK:TIME? reads the 
Time
 field on the 
Marker
 menu of 
the SCOPE screen.