GPIB Command Dictionary
OSCilloscope subsystem
:TRIGger:TYPE <string>
:TRIGger:TYPE?
:TRIGger:TYPE?
These commands set/query how the trigger level is set. ‘Auto’ triggers a
measurement if a triggering signal is not detected within
approximately 50 ms of the last trigger. ‘Norm’ requires a specific
triggering signal before triggering.
measurement if a triggering signal is not detected within
approximately 50 ms of the last trigger. ‘Norm’ requires a specific
triggering signal before triggering.
Syntax
OSC:TRIG:TYPE ‘Auto’
‘Norm’
OSC:TRIG:TYPE?
Screen/field equivalent
OSC:TRIG:TYPE controls the
Auto/Norm
field on the
Trigger
menu of
the SCOPE screen.