Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
316
Chapter 1
 
GPIB Command Dictionary
SANalyzer subsystem
:MARKer:DELTa:NPEak
This command sets the delta marker at the next peak of the spectrum 
analyzer display.
Syntax
SAN:MARK:DELT:NPE
Screen/field equivalent
SAN:MARK:DELT:NPE controls the 
Marker To, Next Peak
 field on 
the 
Marker
 menu of the SPEC ANL screen when the 
Norm/Delta
 field 
is set to 
Delta
.
:MARKer:DELTa:NPLevel <real number>
:MARKer:DELTa:NPLevel?
These commands set/query the level of the next peak signal at the delta 
marker.
This command utilizes the “Real Number Setting Syntax” on page  299, 
but does not use the :STATe command.
Syntax
SAN:MARK:DELT:NPL <real number>
SAN:MARK:DELT:NPL?
Screen/field equivalent
SAN:MARK:DELT:NPL controls the 
Level
 field on the SPEC ANL 
screen when the 
Norm/Delta
 field is set to 
Delta
.
:MARKer:DELTa:PEAK
This command moves the delta marker to the highest peak of the 
signals on the spectrum analyzer display trace.
Syntax
SAN:MARK:DELT:PEAK
Screen/field equivalent
SAN:MARK:DELT:PEAK controls the 
Marker To
Peak
 field on the 
Marker
 menu of the SPEC ANL screen when the 
Norm/Delta
 field is set 
to 
Delta
.