Agilent Technologies E6380-90073 Manuale Utente

Pagina di 352
Chapter 1
321
GPIB Command Dictionary
SANalyzer subsystem
:MASK:BEEP
:MASK:BEEP?
This command turns the failure indicator beeper off or on. 
Syntax
SAN:MASK:BEEP ‘On’
              ‘Off’
SAN:MASK:BEEP?
Screen/field equivalent
SAN:MASK:BEEP controls the 
Mask Beep
 field on the 
Mask
 menu of 
the SPEC ANL screen.
:MASK:DISPlay
:MASK:DISPlay?
This command sets/queries which masks will be displayed: upper, 
lower, both, or none (Off). 
Syntax
SAN:MASK:DISP ‘UpperOnly’
              ‘LowerOnly’
              ‘Both’
‘Off’
SAN:MASK:DISP?
Screen/field equivalent
SAN:MASK:DISP controls the 
Display
 field on the 
Mask
 menu of the 
SPEC ANL screen.
:MASK:FIXed:LOWer:POINts:NUMBer
:MASK:FIXed:LOWer:POINts:NUMBer?
This command sets/queries the number of points in the lower spectrum 
analyzer mask. The maximum number is 15. 
Syntax
SAN:MASK:FIX:LOW:POIN:NUMB <integer 1-15>
SAN:MASK:FIX:LOW:POIN:NUMB?
Screen/field equivalent
SAN:MASK:FIX:LOW:POIN:NUMB controls the 
#Pts
 field on the 
Mask
 
menu of the SPEC ANL screen when the 
Edit Mask
 field is set to 
Lower
, and the 
Mask Type
 field is set to 
Fix
.