Chapter 1
61
GPIB Command Dictionary
AFANalyzer subsystem
:NOTCh:GAIN <string>
:NOTCh:GAIN?
These commands set/query the gain of the AF analyzer’s notch filter 
amplifier. This command is only effective when autoranging is turned 
off using the AFAN:RANG ‘Hold’ command. 
Gain is automatically set when the AFAN:RANG ‘Auto’ command is 
used. If autoranging is enabled (AFAN:RANG ‘Auto’) and you attempt 
to change the gain setting by sending the AFAN:NOTC:GAIN 
command, the value you send will be overridden by the autoranging 
function.
Syntax
AFAN:NOTC:GAIN ‘0 dB’
 ‘10 
dB’
 ‘20 
dB’
 ‘30 
dB’
 ‘40 
dB’
AFAN:NOTC:GAIN?
Screen/field equivalent
AFAN:NOTC:GAIN controls the 
Notch Gain
 field on the AF 
ANALYZER screen when the 
Gain Cntl
 field is set to 
Hold
.
:NOTCh:FREQuency <real number>
:NOTCh:FREQuency?
These commands set/query the center frequency for the variable 
frequency notch filter. This command utilizes the “Real Number Setting 
Syntax” on page 299, but
 does not use the :STATe command.
Syntax
AFAN:NOTC:FREQ <real number> !valid 330 Hz to 10 kHz
AFAN:NOTC:FREQ?
Screen/field equivalent
AFAN:NOTC:FREQ controls the 
Notch Freq
 field on the AF 
ANALYZER screen.