Agilent Technologies E6380-90073 Manuale Utente
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
Chapter 1
67
GPIB Command Dictionary
AFG2|ENCoder subsystem
AFG2|ENCoder subsystem
:AM <real number>
:AM?
:AM?
This command sets AM modulation depth if the command AFG2:DEST
‘AM’ is used first.
‘AM’ is used first.
This command utilizes the “Real Number Setting Syntax” on page 299.
Syntax
AFG2:AM <real number>
AFG2:AM? !returns the value of the amplitude modulation
Screen/field equivalent
AFG2:AM controls the
AFGen2 To
field (lower subfield) on the RF
GENERATOR screen when the upper subfield is set to
AM
.
:BURSt <integer>
:BURSt?
:BURSt?
These commands set/query the number of times the message data is
output. To use this function, the AFG2:SEND:MODE command must
specify ‘Burst’ first.
output. To use this function, the AFG2:SEND:MODE command must
specify ‘Burst’ first.
This command uses only the :INCR command of the “Integer Number
Setting Syntax” on page 206.
Setting Syntax” on page 206.
Syntax
AFG2:BURS <integer>
AFG2:BURS?
Screen/field equivalent
AFG2:BURS controls the
Bursts
field on the SIGNALING ENCODER
screen when the
Mode
field is set to
DTMF
,
AMPS-TACS
, or
NAMP-NTAC
.