Agilent Technologies DC140 Manuale Utente

Pagina di 66
 
 
User Manual: Family of 8-bit Digitizers 
 
Page 37 of 66 
3.6. 
Internal Calibration 
The software drivers supplied include calibration functions for the timing, gain and offset settings, which can be 
executed upon user request. The digitizers are never calibrated in an “automatic” way, i.e. as a side effect of another 
operation. This ensures programmers have full control of all calibrations performed through software in order to 
maintain proper event synchronization within automated test applications. 
The model DC and DP series digitizers include a high precision voltage source and a 16-bit DAC, used to determine 
the input voltage and offset calibration.  
For accurate time and voltage measurements it is recommended to perform a calibration once the module has attained 
a stable operating temperature (usually reached with a few minutes of digitizer operation after power on). Further 
calibration should not be necessary unless temperature variations occur. 
A full internal calibration of a digitizer can be very time consuming (> 100 s/digitizer), in particular for the DC2x1A 
models. Therefore, several other options are available. They are documented in the Programmer's Reference 
Manual
. A program can always be started with the digitizer in an uncalibrated state and data taken can be used for 
many kinds of testing. However, as soon as good data respecting the specifications of the instrument is required a 
calibration of at least the current acquisition state is needed. The full internal calibration has the advantage that it 
generates the calibration constants needed for any possible configuration of the instrument; its disadvantage is the 
time taken. If a more selective calibration is done it will allow the generation of good data in the current acquisition 
state. This calibration will remain useable whenever that acquisition state is used again for as long as the temperature 
of the instrument does not change significantly. A fast calibration of a channel in a configuration can be done in 
around a second. Many applications can save time by only performing calibration for the configurations that will 
actually be used. 
3.7. 
ASBus 
The DC Series digitizers may be used in applications that require many data acquisition channels. In such cases it is 
possible to use more than one digitizer in a standard CompactPCI/PXI crate. Each DC Series digitizer, except the 
DC135 and DC140 models, includes ASBus, a proprietary high bandwidth auto-synchronous bus system that allows 
multiple digitizers to work together synchronously.  
The ASBus distributes both the clock and trigger signals along a plug-in front panel bus, between all the digitizers 
that participate in the system. It allows any digitizer to act as the trigger source and any other digitizer to act as the 
clock source (acquisition master), enabling all the digitizers to be clocked at the same time. Synchronizing the clock 
signals between the devices improves the accuracy of cross-channel measurements and is essential for accurate time 
correlation between signals on different channels. 
The ASBus is intended to connect modules of the same type, i.e. of the same model number, although some 
exceptions to this rule might be possible. If modules with the same model number, but different memory length 
options, are connected only the shortest memory length can be used. 
The ASBus connector is located on the front panel of each cPCI module. Bridges are used to connect adjacent 
modules for synchronization, which is then activated through software by defining a master module and the trigger 
source. Up to 7 modules may be synchronized with the ASBus. More detailed information and the commands 
required to setup the ASBus clock and trigger distribution are included in the Programmer's Guide and Reference 
manual. 
The DP1400 uses the ASBus
system with a connector on the top of each card to achieve the same functionality. It 
works for up to 3 modules. 
3.8. 
Special Front Panel Input and Output(DP1400 & DC271-FAMILY) 
 
Connector Function 
CLK IN 
50 Ω External Reference or External Clock Input 
I/O A 
User configurable (see below)  
TR OUT 
Signal occurs after an accepted TRIGGER.  
I/O B 
User configurable (see below) 
The DC271-Family uses MMCX connectors for these functions; the DP1400 uses the MCX type. The I/O A and I/O 
B signals are 3.3 V compatible CMOS. This means that, on input, low is < 0.7 V and high must be in the range [1.7 
V, 5.0 V]. An unconnected signal will be high. This definition ensures TTL compatibility. On output, the low level