Calibration Commands
 
Calibration commands let you enable and disable the calibration mode, change the calibration password,
calibrate current and voltage programming, and store new calibration constants in nonvolatile memory.
NOTE:
If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error. You must also save any changes that you made
using CALibrate:SAVE, otherwise all changes will be lost when you exit calibration mode.
 
CALibrate:CURRent
 
This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
   
 
Command Syntax
 
CALibrate:CURRent[:SOURce][:DC][:POSitive]
 
Parameters
 
None
 
Examples
 
CAL:STAT 1
,
! enable calibration
   
CAL:CURR
,
! start current calibration
 
Related Commands
 
CAL:CURR:NEG,
CAL:LEV,
CAL:DATA
CALibrate:CURRent2
 
Agilent 66309B/D only
This command initiates the current calibration of output 2.
Command Syntax
CALibrate:CURRent2
Parameters
None
Examples
CAL:CURR2
 
Related Commands
 
CAL:CURR:NEG,
CAL:LEV,
CAL:DATA
 
CALibrate:CURRent:MEASure:LOWRange
 
Agilent 66311B/D, 66309B/D only
 
This command initiates the calibration of the low-range current measurement circuit.
   
 
Command Syntax
 
CALibrate:CURRent:MEASure[:DC]:LOWRange
 
Parameters
 
None
 
Examples
 
CAL:CURR:MEAS
 
Related Commands
 
CAL:CURR
 
CALibrate:CURRent:MEASure:AC
 
Agilent 66311B/D, 66309B/D only
 
This command initiates the calibration of the high bandwidth (ac) measurement circuit.
   
 
Command Syntax
 
CALibrate:CURRent:MEASure:AC
 
Parameters
 
None
 
Examples
 
CAL:CURR:MEAS:AC