Agilent Technologies 66309B Manuale Utente

Pagina di 190
 
Language Dictionary - 8
141
Additional Commands
The following commands are added to the language dictionary to provide limited Fluke PM2811 Power
Supply compatibility. These comands will only work when the SYSTem LANGuage is set to SCPI.
INSTrument:STATe
 
This command enables or disables the dc source output. The state of a disabled output is a condition of
zero output voltage and a model-dependent minimum source current (see *RST).
   
 
Command Syntax
 
INSTrument:STATe <bool>
 
Parameters
 
0 | OFF | 1 | ON
 
*RST Value
 
0
 
Examples
 
INST:STAT 1
 
Query Syntax
 
INSTrument[:STATe]?
 
Returned Parameters
 
<NR1> 0 or 1
 
Equivalent SCPI Command
 
OUTPut[:STATe]
OUTPut:PROTection:TRIPped?
 
This query reports if any  OV (overvoltage), OCP (overcurrent protection), OT (over temperature), or RI
(remote inhibit) status bit is set. 0 = bit is not set; 1 = bit is set.
   
 
Query Syntax
 
OUTPut:PROTection:TRIPped?
 
Returned Parameters
 
<NR1> 0 or 1
 
Examples
 
OUTP:PROT:TRIP?
 
Equivalent SCPI Command
 
STATus:QUEStionable:CONDition?
CURRent:LIMit:HIGH?
 
This query returns the maximum rated current of the instrument.
   
 
Query Syntax
 
[SOURce:]CURRent:LIMit:HIGH?
 
Returned Parameters
 
<NRf>
 
Examples
 
CURR:LIM:HIGH?
 
Equivalent SCPI Command
 
[SOURce:]CURRent? MAX
CURRent:LIMit:LOW?
 
This query returns the minimum rated current of the instrument.
   
 
Query Syntax
 
[SOURce:]CURRent:LIMit:LOW?
 
Returned Parameters
 
<NRf>
 
Examples
 
CURR:LIM:LOW?
 
Equivalent SCPI Command
 
[SOURce:]CURRent? MIN