7 - Programming the DC Source
84
 
Triggered Measurements
 
Use the measurement trigger system to synchronize the acquisition of measurements with either a BUS or
internal trigger. You can trigger voltage and current measurements on the main output (output 1) and on
the DVM. An internal trigger synchronizes the acquisition to a signal condition. Use  FETCh commands
to return different calculations from the data acquired by the measurement system. Briefly, to make a
triggered measurement:
1
Select a sweep interval and sample size
2
Set up the trigger levels
3
Setting the output 2 voltage and current
4
Select the trigger source
5
Enable the trigger system
6
Fetch the triggered measurements
 
SCPI Triggering Nomenclature
 
The dc source uses the following sequence name and alias for the measurement trigger system. The alias
can be used instead of the sequence form.
 
Sequence Form
 
Alias
 
SEQuence2
 
ACQuire
 
Measurement Trigger Model
 
Figure 7-4 is a model of the measurement trigger system. The rectangular boxes represent states. The
arrows show the transitions between states.  These are labeled with the input or event that causes the
transition to occur.
 
ACQUIRED
INITIATED STATE
IDLE STATE
ABORt
*RCL
*RST
TRIGGER RECEIVED
INITiate[:IMMediate]
TRIGger:COUNt
COMPLETE?
YES
NO
SENSe:SWEep:POINts
 Figure 7-4. Model of Measurement Trigger System