Agilent Technologies FS2101 Manuale Utente

Pagina di 48
RESRVD 0100 
RESRVD 0101 
MEM_RD 0110 
MEM_WR 0111 
RESRVD 1000 
RESRVD 1001 
CON_RD 1010 
CON_WR 1011 
MEMRDM 1100 
DAC_CY 1101 
MEMRDL 1110 
MEMWRI 1111 
I/O_XACTIONS 001X 
MEM_XACTIONS 011X 
CONFIG_XACTIONS 101X 
 
 
 
 Theory of Operation
 
The PCI Active Analysis Probe is a universal PCI short card that 
attaches to  Agilent logic analyzers.  The Analysis Probe has five 
major parts: 
1.  The input buffers 
2.  The latching buffers 
3.  The interface to the logic analyzer 
4.  The clocking and cycle bit generation logic 
5.  The extender card connector 
 
 
The Input Buffers 
The input buffers present a single electrical load on the PCI bus 
and are made up of IDT162260 tri-port buffers.  The PCI clock is 
buffered by a Motorola 807 high speed clock buffer.   
When the Analysis Probe is in Timing mode output Port 2 of the 
tri-port buffers goes directly to the logic analyzer input 
terminators.  When in State mode output Port 2 is tri-stated.   
Output port 1 is always enabled and goes to the latching buffers.
 