How to clean connector adapters
Cleaning Information
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
181
To assess the projection of the emitted light beam you can use an 
infrared sensor card. Hold the card approximately 5 cm from the 
output of the connector. The invisible emitted light is projected onto 
the card and becomes visible as a small circular spot.
Preferred Procedure
Use the following procedure on most occasions.
Clean the connector by rubbing a new, dry cotton swab over the 
surface using a small circular movement.
Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the connector:
Moisten a new cotton swab with isopropyl alcohol.
Clean the connector by rubbing the cotton swab over the surface 
using a small circular movement.
Take a new, dry soft tissue and remove the alcohol, dissolved 
sediment and dust, by rubbing gently over the surface using a small 
circular movement.
Blow away any remaining lint with compressed air.
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use 
an ultrasonic bath with isopropyl alcohol.
Hold the tip of the connector in the bath for at least three minutes.
Take a new, dry soft tissue and remove the alcohol, dissolved 
sediment and dust, by rubbing gently over the surface using a small 
circular movement.
Blow away any remaining lint with compressed air.
How to clean connector adapters
C A U T I O N
Some adapters have an anti-reflection coating on the back to reduce 
back reflection. This coating is extremely sensitive to solvents and 
mechanical abrasion. Extra care is needed when cleaning these 
adapters.