How to clean instruments with an optical glass plate
Cleaning Information
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
185
If there are fluids or fat in the connector, please refer the instrument 
to the skilled personnel of Agilent’s service team.
C A U T I O N
Only use clean, dry compressed air. Make sure that the air is free of 
dust, water, and oil. If the air that you use is not clean and dry, this 
can lead to filmy deposits or scratches on the surface of your 
connector interface. This will degrade the performance of your 
transmission system. 
Never try to open the instrument and clean the optical block by 
yourself, because it is easy to scratch optical components, and cause 
them to become misaligned.
How to clean instruments with 
an optical glass plate
Some instruments, for example, the optical heads from Agilent 
Technologies have an optical glass plate to protect the sensor. Clean 
this glass plate in the same way as optical lenses (see “How to clean 
How to clean instruments with a 
physical contact interface
Remove any connector interfaces from the optical output of the 
instrument before you begin the cleaning procedure.
Cleaning interfaces is difficult as the core diameter of a single-mode 
fiber is only about 9
µm. This generally means you cannot see streaks 
or scratches on the surface. To be certain of the degree of pollution on 
the surface of your interface and to check whether it has been 
removed after cleaning, you need a microscope.