Agilent Technologies 81642A Manuale Utente

Pagina di 194
Specifications
Definition of Terms
44
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, 
Sixth Edition
Power Flatness Over Modulation
When changing the wavelength and modulation frequency, and 
measuring the differences between actual and displayed power levels 
(in dB), the power flatness is ± half the span between the maximum 
and the minimum value of all differences.
Conditions: uninterrupted line voltage, constant power setting, 
temperature within ±2 K, external modulation ON.
Measurement with optical power meter.
Power Flatness Versus Wavelength
When changing the wavelength at constant power setting and 
recording the differences between actual and displayed power levels, 
the power flatness is 
± half the span (in dB) between the maximum 
and the minimum of the measured power levels.
Conditions:  uninterrupted TLS output power, constant power setting, 
temperature within 
±1K.
Measurement with optical power meter.
Power Linearity
When changing the power level and measuring the differences (in dB) 
between actual and displayed power levels, the power linearity is 
± 
half the span (in dB) between the maximum and the minimum value of 
all differences.
Conditions:  power levels from within specified output power range, 
uninterrupted TLS output power, at fixed wavelength settings and 
stable temperature.  
Measurement with optical power meter.
Power Repeatability
The random uncertainty in reproducing the power level after changing 
and re-setting the power level.  The power repeatability is 
± half the 
span (in dB) between the highest and lowest actual power.
Conditions:  uninterrupted TLS output power, constant wavelength, 
temperature within 
±1 K, short time span.  
Measurement with optical power meter.
N O T E
The long-term power repeatability can be obtained by taken the power 
repeatability and power stability into account.