Performance Tests
Performance Test Instructions
62
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, 
Sixth Edition
General Settings of Wavelength Meters for all 
Wavelength Tests
Set the Burleigh WA-150 to the following settings:
• Set Display to Wavelength.
• Set Medium to Vacuum.
• Set Resolution to Auto.
• Set Averaging to On.
• Set Input Attenuator to Auto. 
Wavelength Accuracy
The steps below explain how to calculate the Relative Wavelength 
Accuracy, Absolute Wavelength Accuracy, and the Mode Hop Free 
Tuning Result.
Relative Wavelength Accuracy
Move to the Tunable Laser channel of the Agilent 8164A Lightwave 
Measurement System and press [Menu].
Set the menu parameters to the values shown in Table Table 5. 
If you use the Agilent 81480A, Agilent 81680A Tunable Laser 
module or the Agilent 81640A Tunable Laser module:
Connect the fiber output to Output 2, the High Power output.
Set <Optical Output> to <High Power (2)>.
Set the wavelength and power of your Tunable Laser module to the 
values given in Table Table 6.
Table 5
Tunable Laser Channel Settings
Tunable Laser Channel Menu Parame-
ters
Values
<Wavelength Mode>
<
λ>
<Source State>
<Off>
<Power Unit>
<dBm>
<Power Mode>
<Automatic>