Performance Test Instructions
Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
67
Power Tests
Calibration of the Agilent 81001FF Attenuation Filter
N O T E
When an HP 81524A Optical Head is used in conjunction with a 
Agilent 81001FF Attenuation Filter, it is absolutely necessary that you 
calibrate the Agilent 81001FF Attenuation Filter before starting the 
power tests in the following measurement setups; it is not sufficient to 
use calibration factors that are derived from an earlier setup.
Make sure all instruments have warmed up. 
Figure 18
Calibration of the Agilent 81001FF Attenuation Filter, Reference Setup
Set the power meter installed in the HP 8153A to the following 
values:
a
Set range to 0 dBm; press Up or Down as required.
b
Set T, the averaging time, to 500 ms.
c
Agilent 81680A, 40A, 82A, 42A, & 89A: 
Set 
λ, the wavelength, to 1550 nm.
Agilent 81480A:
Set 
λ, the wavelength, to 1450 nm.
d
Select dBm as the power units.
Move to the Tunable Laser channel of the Agilent 8164A.
Agilent 81680A, 40A, 82A, 42A, & 89A:
Set 
λ, the wavelength, to 1550 nm, and [P], the power, to 0 dBm..
For Agilent 81480A:
Set 
λ, the wavelength, to 1450 nm, and [P], the power, to 0 dBm.
Press the key beside the laser output to switch the laser on.
Agilent 8164A Lightwave
Measurement System
HP 81524A
Tunable Laser
For #021, #071: use 81000FI and 81113PC
For #022, #072: use 81000SI and 81113SC
Slanted 
Connector
Optical Head
Agilent 81000SA
HP 81533B
HP 8153A Lightwave Multimeter