Performance Test Instructions
Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
71
Power Linearity
Power Linearity - Low Power Test
To measure the power linearity of the Low SSE output, Output 1, of 
the Agilent 81480A, Agilent 81680A, or the of the Agilent 81640A:
Set up the equipment as shown in Figure 21
Figure 21
Test Setup for Low Power Linearity Tests
Move to the Tunable Laser channel of the Agilent 8164A Lightwave 
Measurement System and press [Menu].
Set the menu parameters to the values shown in Table Table 5. 
<Power Mode> does not apply.
Set the wavelength and power for each Tunable Laser module to the 
values given in Table Table 11.
If you are using the Agilent 81680A, Agilent 81480A, or the 
Agilent 81640A Tunable Laser module: 
Connect the output fiber to Output 1, the Low SSE output.
Set <Optical Output> to <Low SSE (1)>.
Make sure the optical output is switched off.
Agilent 8164A Lightwave
Measurement System
Tunable Laser
For #021, #071: use 81000FI and 81113PC
For #022, #072: use 81000SI and 81113SC
Slanted 
Connector
HP 81532A
HP 8153A Lightwave Multimeter
Power Sensor
Table 11
Wavelength and Power Settings for Low Power Linearity Tests
Module
Wavelength
 [λ] 
Power [
P]
Agilent 81480A - Output 1
1430.000 nm
−7.00 dBm
Agilent 81680A - Output 1
1540.000 nm
−6.00 dBm
Agilent 81640A - Output 1
1560.000 nm
−7.00 dBm