Performance Test Instructions
Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
79
Power Stability
Follow the steps below to measure the power stability:
Set up the equipment as shown in Figure 20.
Move to the Tunable Laser channel of the Agilent 8164A Lightwave 
Measurement System and press [Menu].
Set the menu parameters to the values shown in Table Table 5.
Set the wavelength and power for each Tunable Laser module to the 
values given in Table Table 16.
If you are using the Agilent 81480A, Agilent 81680A or the 
Agilent 81640A Tunable Laser module: 
Connect the output fiber to Output 1, the Low SSE output.
Set <Optical Output> to <Low SSE (1)>.
Ensure the optical output is switched off.
Set the power meter channel of the HP 8153A to the following 
settings:
• Enable automatic ranging; press Auto as required.
• Set T, the averaging time, to 200 ms.
N O T E
You should record measurements after a time interval of at least 3 
seconds.
• Set the 
λ, the wavelength, to the same as your Tunable Laser 
module, as given in Table Table 16.
• Select dB as the power units.
Table 16
Wavelength and Power Settings for Power Stability Tests
Module
Wavelength
 [λ] 
Power [
P]
Agilent 81480A - Output 1
1450.000 nm
−13.000 dBm
Agilent 81480A - Output 2
1450.000 nm
− 3.000 dBm (ATT = 0 dB)
Agilent 81680A - Output 1
1540.000 nm
−13.000 dBm
Agilent 81680A - Output 2
1540.000 nm
− 3.000 dBm (ATT = 0 dB)
Agilent 81640A - Output 1
1560.000 nm
−13.00 dBm
Agilent 81640A - Output 2
1560.000 nm
− 5.00 dBm (ATT = 0 dB)
Agilent 81682A
1540.000 nm
− 3.000 dBm
Agilent 81682A #003
1540.000 nm
− 5.500 dBm (ATT = 0 dB)
Agilent 81642A
1580.000 nm
− 3.000 dBm
Agilent 81642A #003
1580.000 nm
− 4.500 dBm (ATT = 0 dB)
Agilent 81689A
1540.000 nm
− 3.00 dBm