Agilent Technologies 85036B Manuale Utente

Pagina di 49
85036B/E
1-7
General Information
Calibration Kits Documented in this Manual
Calibration Kits Documented in this Manual
This manual applies to any 85036B or 85036E calibration kit whose serial prefix is listed 
on the title page. If your calibration kit has a different serial number prefix than the one 
listed on the title page, refer to 
 below for information on how this 
manual applies.
Calibration Kit History
This section describes calibration kits with serial number prefixes different than those 
listed on the title page.
Agilent 85036B kits with serial prefix 3102A and below
• 75 ohm load specification: return loss 
 45 db, dc to 2 GHz
• all other kit components unspecified
85036B calibration kits with serial number prefix 3102A and below should meet the 
specifications published in this manual. If you wish to have your kit recertified to current 
specifications, contact your nearest Agilent service center. For contact information, see 
.
Agilent 85036B kits with serial prefix 3103A and below
These calibration kits had a center conductor extender with a smaller body. Therefore, a 
different procedure was used for connecting the female open. If you’re using one of the 
smaller extenders, replace steps 1 through 3 in 
 of this manual with the steps below.
1. Mate the center conductor extender to the male pin of the test port.
2. Connect the open body to the test port by carefully guiding it over the extender.
3. To disconnect the open, carefully remove the open body first.
The part numbers in this manual are the recommended replacement parts for these kits 
(including the center conductor extender).