4-2
85036B/E
Performance Verification
Introduction
Introduction
The performance of your calibration kit can only be verified by returning the kit to Agilent 
Technologies for recertification. The equipment required to verify the specifications of the 
devices in the kit has been specially manufactured and is not commercially available.
How Agilent Verifies the Devices in This Kit
Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision 
airlines and shorts that are directly traced to NIST (National Institute of Standards 
and Technology). The airline and short characteristics are developed from mechanical 
measurements. The mechanical measurements and material properties are carefully 
modeled to give very accurate electrical representation. The mechanical measurements 
are then traced to NIST through various plug and ring gages and other mechanical 
measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) 
testing of the calibration devices, Agilent includes the test measurement uncertainty as 
a guardband to guarantee each device meets the published specification. For 
recertifications (after sale), no guardband is used and the measured data is compared 
directly with the specification to determine the pass or fail status. The measurement 
uncertainty for each device is, however, recorded in the calibration report that 
accompanies recertified kits.
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the 
institute’s calibration facility. The specifications data provided for the devices in this kit is 
traceable to NIST through Agilent Technologies.