85036B/E
 
Standard Definitions
Nominal Standard Definitions
Nominal Standard Definitions
Standard definitions provide the constants needed to mathematically model the electrical 
characteristics (delay, attenuation, and impedance) of each calibration standard. The 
nominal values of these constants are theoretically derived from the physical dimensions 
and material of each calibration standard, or from actual measured response. These values 
are used to determine the measurement uncertainties of the network analyzer. The 
standard definitions in
 list typical calibration kit parameters used to specify the 
mathematical model of each device. This information must be loaded into the network 
analyzer to perform valid calibrations. Refer to your network analyzer’s user’s guide for 
instructions on loading calibration constants.
NOTE
The values in the standard definitions table are valid only over the specified 
operating temperature range.
Setting the System Impedance
This kit contains only 75 ohm devices. Ensure the system impedance (Z
0
) is set to 75 ohms. 
Refer to your network analyzer’s user’s guide for instructions on setting system impedance.