9.2.2  P1.2: Select Calibration Buffer Options 
The types of pH calibration buffer options available are:  
Meter 
Meter Display 
pH Calibration Buffer Options 
pH 11 
USA  
nSt 
USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45. 
NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45. 
pH 110  USA 
nSt 
dn 
PUr 
USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45. 
NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.  
DIN buffers --- pH 1.09, 3.06, 4.65, 6.79, 9.23, and 12.75. 
PWB ( Low Ionic): pH 4.10 and 6.97 
From measurement mode 
Press SETUP key to enter Set Up mode.  
1.  Press the MI/▲ or MR/▼ keys to 
scroll through subgroups until you 
view parameter P1.0.  
2.  Press the HOLD/ENTER key two 
times to select parameter 1.2.  
3.  Press the MI/▲ or MR/▼ keys to 
select the buffer set you require: 
4.  Press the HOLD/ENTER key to 
confirm selection and to return to the 
subgroup menu.  Press CAL/MEAS 
key to return to measurement mode.  
See Figure 31.  
 
 
 
 
 