eutech-instruments cyberscan ph 110 phmvrs232 meter Manuale Utente

Pagina di 76
Instruction Manual   
 
CyberScan pH 11 / 110  
 
39
9.4 
P3.0: View Electrode Data 
Program 3 has two “view only” options that let you check the electrode parameters 
for diagnostic purposes.  
• 
From pH measurement mode, you can view the electrode’s offset and slope 
values.  
•  From mV measurement mode, you can view the electrode’s relative mV offset 
value. (pH 110 only) 
9.4.1  View pH Electrode Offset and Slope 
 From pH measurement mode: 
1.  Press the SETUP key to enter Set Up mode.  
2. Press 
the 
MI/▲ or MR/▼ keys to scroll through 
subgroups until you view parameter P3.0.   
3.  Press the HOLD/ENTER key. 
4.  The display shows the electrode offset value.  It 
is the mV offset at pH 7.00. If you have not 
calibrated at any buffer, the primary display 
shows ‘---‘.  
5.  Press the HOLD/ENTER key to proceed to 
electrode slope display.  
6. The display shows electrode slope in 
percentage. Slope displayed will be the current 
slope of the slope zone to which a measurement 
is made or calibration is done. 
7.  At any point, you can press the CAL/MEAS key 
to return to measurement mode.  
See Figure 35.  
 
 
 
Figure 35: P3.1/3.2 - View pH 
electrode offset and slope
mV
pH
SETUP
5.3
 p3.1
SETUP
 p 3.0
ele
%
pH
SETUP
 p 3.2
100.0
ENTER
HOLD
ENTER
HOLD