SPECIFICATIONS:
RANGE
RANGE
0.000 1.999 µS/cm
RESOLUTION
0.001 µS/cm
ACCURACY
(@20°C/68°F)
±2% f.s.
TYPICAL EMC DEVIATION ±0.5% f.s.
ENVIRONMENT
ENVIRONMENT
0 to 50 C (32 to 122 F);
95% RH not condensing
BATTERY TYPE
4 x 1.5V alkaline (included)
BATTERY LIFE
approx. 120 hours
of continuous use
DIMENSIONS
175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT
78 g (2.7 oz.)
IS98309-R4 07/01
ELECTRODE REPLACEMENT:
The electrode is built-in inside the plastic casing and
can be easily replaced in the following way:
can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the UPW
UPW
UPW
UPW
UPW
located below the battery compartment and slide
off the whole internal part of the tester.
off the whole internal part of the tester.
• Replace the built-in electrode with a new
HI73309.
• Before reinserting the internal circuit into the
new plastic casing, make sure the O-rings (2)
of the metal tube (3) are properly seated (see
drawing).
To avoid bending the connections to the con-
ductivity sensors (1), do not apply excessive
force while inserting the circuit. In case of any
problem, reslide and reinsert it.
of the metal tube (3) are properly seated (see
drawing).
To avoid bending the connections to the con-
ductivity sensors (1), do not apply excessive
force while inserting the circuit. In case of any
problem, reslide and reinsert it.
Internal circuit:
ACCESSORIES:
HI 73309
Spare electrode
HI 731326
Calibration screwdriver (20 pcs)
• Fasten the screw on the back, below the
battery compartment.
• Reinsert the battery compartment cover.