Instruction Manual 
 
Ammonia Gas-sensing Electrode
 
 
18
                                                 
 
 
 
value correctly 
 
                          
incorrect standards       
 
prepare fresh standards 
 
                          
wrong units used          
 
apply correct conversion factor: 
                                                 
 
 
 
10-
3
M = 17 ppm as 
                                                 
 
 
 
NH
3
 = 14 ppm as N 
                                                 
                          
buffer added to           
 
add same proportions of buffer to             
                         
standards and not                        
standards and samples 
   samples 
 
 
                          
sample carryover         
 
rinse electrodes thoroughly 
                                                 
 
 
 
between samples 
 
 
Checking the Electrode Inner Body 
 
If the electrode slope is found to be low during operation, the following solutions will be necessary 
to check the inner body: 
 
- pH 4 Buffer (0.1M NaCl added) 
Add 2.9 grams of reagent-grade NaCl to 500 ml of pH 4 buffer. Dissolve the solid. The solution 
may be stored for repeated use. 
 
- pH 7 Buffer (0.1M NaCl added) 
Add 2.9 grams of reagent-grade NaCl to 500 ml of pH 7 buffer. Dissolve the solid. The solution 
may be stored for repeated use. 
 
Disassemble the ammonia electrode. If the electrode is dry, soak the glass tip of the inner body in 
Eutech Ammonia Electrode Filling Solution (Code No. EC-RE-032) for at least two hours. 
 
Rinse the electrode thoroughly with distilled water. Put 100 ml of pH 7 buffer (0.1M NaCl added) 
in a 150 ml beaker. Place the beaker on the magnetic stirrer, and begin stirring. Immerse the tip of 
the inner body in the solution so that the reference element is covered. Make sure that the meter is 
in the mV mode. Record the meter reading when stable. 
 
Rinse the inner body thoroughly in distilled water. Put 100 ml of pH 4 buffer (0.1M NaCl added) in 
a 150 ml beaker, place the beaker on the magnetic stirrer, and begin stirring. Immerse the tip of the 
inner body in the solution so that the reference element is covered. Observe the change in the meter 
reading carefully. In less than 30 seconds after immersion, the reading should change 100 mV. The 
meter reading should stabilize in 3 - 4 minutes, with a difference greater than 150 mV if the inner 
body sensing elements are operating properly.