Kern FFN 15K2IPNParcel scales Weight range bis 15 kg FFN 15K2IPN Manuale Utente
![Kern](https://files.manualsbrain.com/attachments/04ab1b8985747ba02c8b0a2770613923db7d4965/common/fit/150/50/4233fd5bdc88993dc151edc4fd729c0efa33c251fd07d340e43459f4e2b9/brand_logo.gif)
Codici prodotto
FFN 15K2IPN
19
FFN-N-BA-e-1222
7.1.1 Adjusting non-verifiable models
Display
Operation
Start balance by pressing
Whilst the weighing scale is carrying out a self-test,
(00…- 99…) press
and hold until the message “F1
CAL“ appears.
Press
,
in the display appears „UnLod“.
(Example)
Press
again
the weight value will be displayed.
Enter value for adjustment weight (See chpt 1) by
Position weight
Press
Remove weight whilst scale is carrying out self-test
Weighing scale changes to zero display.
This completes the adjustment process.
Weighing scale changes to zero display.
This completes the adjustment process.