Kern Precision scales PCB 250-3 Weight range 250 g Readability 0.001 g mains-powered, battery-powered, rechargeable Silv PCB 250-3 Scheda Tecnica
Codici prodotto
PCB 250-3
PCB-BA-e-1315
18
Back to weighing mode by
pressing the MODE button.
Error message „Er 1“
Piece below minimum weight of piece (See chpt. 1
“Technical specifications“): Press MODE key and restart
reference determination.
Piece below minimum weight of piece (See chpt. 1
“Technical specifications“): Press MODE key and restart
reference determination.
Taring
The tare vessels can also be used for piece counting.
Before starting piece counting use the TARE
button to tare out the container.
Before starting piece counting use the TARE
button to tare out the container.
Net-total
weighings
weighings
SET
M
It is useful if a mixture of several components is weighed into a
tare vessel and finally the sum weight of all weighed
components is necessary for control purposes (net-total, i.e.
the weight of the tare vessel).
Example:
tare vessel and finally the sum weight of all weighed
components is necessary for control purposes (net-total, i.e.
the weight of the tare vessel).
Example:
1. Place tare container on the weighing plate. Press the TARE
button, the zero display disappears.
2. Weigh-in component . Press the SET button, the zero
display disappears. [] is displayed on the right border of
the display.
the display.
3. Weigh-in component and press SET key. Net-total (sum
weight of the components and ) is displayed.
4. Press the SET button, the zero display disappears.
5. Weigh-in component and press SET key. Net-total (sum
weight of the components and and .) is displayed.
If necessary, also fill the formula up to the desired final
value. For every component more repeat the steps 4-5.
Back to weighing mode by pressing the TARE button.