Kern Precision scales PCB 6000-0 Weight range 6 kg Readability 1 g mains-powered, rechargeable Silver PCB 6000-0 Scheda Tecnica

Codici prodotto
PCB 6000-0
Pagina di 37
PCB-BA-e-1315 
16 
 
 
  Weigh the material, the net weight will be indicated. 
 
The taring process can be repeated any number of times, 
e.g. when adding several components for a mixture 
(adding). The limit is reached when the whole weighing 
range is exhausted. 
The weight of the weighing container will be displayed as a 
minus number after removing the weighing container. 
The tare weight is saved until it is deleted.  
Delete tare 
TARE
 
 Unload the balance and press the TARE button, the zero 
display appears. 
 
 
 
 
 
PRE-TARE-
Function 
MODE
CAL
 
Using this function the weight of a tare vessel is stored. Even 
after turning off/on the weighing balance will continue working 
with the saved tare value. 
 In weighing mode put tare vessel on the weighing plate 
 Press repeatedly the MODE 
key until „PtArE“ flashing 
appears. 
 Use SET key to store the current weight on the weighing 
plate as a PRE-TARE value. 
Delete PRE-TARE 
value 
MODE
CAL
 
 Remove all loads from the balance and press repeatedly 
the MODE 
key until „PtArE“ flashing appears. 
 Use the SET key to confirm. The PRE-TARE value is 
deleted, the zero display appears.