1.1.6 
1 - 3 
Scanning mode: 
TV/FAST 
 Slow 
 
Reduced area (320 x 240 pixels) 
 Waveform 
display 
 
 
Split screen/dual mag mode 
 
High-definition capture (5120 x 3840 pixels maximum) 
 
Frame integration capture (2560 x 1920 pixels maximum) 
 
Scanning speed: 
TV scanning (640 x 480 pixel display: 25/30* frames/s) 
 
FAST scanning (full screen display: 6.3/7.5* frames/s) 
 Slow 
scanning 
 
(full screen display: 1, 4, 20/16, 40/32, 80/64* s/frame)   
 
(640 x 480 pixel display
:  0.5, 2, 10/8, 20/16, 40/32* s/frame) 
 
Reduced area scanning (320 x 240 pixel display: 25/30, 3/4* 
frames/s) 
 
High-definition capture scanning 
 
(2560 x 1920 pixels: 40/32, 80/64, 160/128* s/frame) 
 
(5120 x 3840 pixels: 80/64, 160/128, 320/256* s/frame) 
 
* Power supply synchronization: 50/60 Hz 
 
Auto-tuning feature: 
Auto brightness & contrast control (ABCC) 
 
Auto focus control (AFC) 
 
Auto stigma & focus (ASF) 
 
Auto filament saturation (AFS) 
 
Auto beam alignment (ABA) 
 
Auto beam setting (ABS) 
 
(Auto filament saturation + auto beam alignment + auto gun 
bias alignment) 
 
Auto axial alignment (AAA) 
 
(applicable to aperture/stigma alignment) 
 
Signal/image processing function:  S/N improvement by averaging (TV/FAST scanning) 
 
S/N improvement by pixel integration 
 
Frame integration (maximum number of integrated frames: 
1024) (TV, FAST, SLOW1 scanning) 
 
2-color synthesized image display (realtime/saved image in   
2-image simultaneous display) 
 
Pseudo-color image display (saved image) 
 
Realtime image processing (brightness reversal, gamma 
adjustment) 
 
Realtime histogram display 
 
Image processing on saved images 
 
(brightness reversal, gamma adjustment, various space filter 
processing)