Fluke RUSKA 7350 Manuale Utente

Pagina di 78
RUSKA 7350 
Users Manual 
B-2 
Value 
Description and Corrective Action 
-286 
Program Runtime Error. An error occurred while running the program. Usually the 
setpoint is out of range. 
-313 
Calibration Data Lost. The calibration data has been lost and the unit must be 
recalibrated. 
-315 
Configuration Data Lost. The configuration data has been lost. Check all parameters 
to be sure that they are correct. 
-330 
Self Test failed. See display for more information. 
-350 
Queue Overflow. The error queue was full and messages were lost. 
-400 
Query Error. A read request was received when there was nothing to read. 
500 
Controller Malfunction. Internal control failure. 
501 
High Limit Exceeded. The pressure was greater than the high limit. 
502 
Low Limit Exceeded. The pressure was less than the low limit. 
503 
Slew Limit Exceeded. The pressure changed faster than the slew limit allowed. 
521 
Pressure Overrange. The pressure reading is outside the range of the RUSKA 7350.
538 
Auto vent. The Vent Limit was exceeded and the instrument entered Vent Mode. 
543 
Supply Pressure. The instrument is attempting to increase pressure and the 
pressure is not responding. Check main pressure supply, shop air supply, and for 
leaks. 
600 
Factory Data Lost.  Internal factory constants have been lost. Contact Fluke for more 
information. 
602 
Model Changed. Firmware has been downloaded that is not correct. Download the 
correct firmware. 
706 
Pump position failure. The position sensor is not detecting movement in the pump 
after the pump has been commanded to move. 
707 
Range Switch. The current range can only be changed when the pressure reading is 
below 100 psi (700 kPa).