Agilent E1441A Function/Arbitrary Waveform Generator Module Setup  
Chapter 1
Setting the Module Address Switch
The logical address switch factory setting is 80. Valid addresses are from 1 
to 254 for static configuration (the address you set on the switch) and 
address 255 for dynamic configuration. The Agilent E1441A supports 
dynamic configuration of the address. This means the address is set 
program- matically by the resource manager when it encounters a module 
with address 255 that supports dynamic configuration.
If you install more than one Function Generator, each module must have a 
different logical address. If you use a VXIbus command module, the logical 
address must be a multiple of eight (e.g., 80, 88, 96, etc.) Each instrument 
must have a unique secondary address which is the logical address divided 
by eight.
Note
When using an Agilent E1405A/B or E1406A as the VXIbus resource 
manager with SCPI commands, the Function Generator's address switch 
value must be a multiple of 8. 
Figure 1-1. Setting the Logical Address