Agilent E1441A Error Messages  
141
Appendix B
Appendix B
Agilent E1441A Error Messages
Errors are retrieved in first-in-first-out (
FIFO
) order. The first error returned 
(SYST:ERR? command) is the first error that was stored. When you have 
read all errors from the queue, the next execution of  SYST:ERR? returns 
+0, "No error". Since no indication is given that an operation has caused an 
error, your application program should query the error queue regularly.
If more than 20 errors have occurred, the last error stored in the queue (the 
most recent error) is replaced with -350, “Too many errors”. No additional 
errors are stored until you remove errors from the queue. If no errors have 
occurred when you read the error queue, the function generator responds 
with +0, “No error”.
The error queue is cleared when power has been off or after a *CLS (clear 
status) command has been executed. The *RST (reset) command does not 
clear the error queue. 
SYSTem:ERRor?
Reads one error from the error queue
Errors have the following format (the error string may contain 
up to 80 characters):
-113,"Undefined header"
Execution Errors
-101
Invalid character 
An invalid character was found in the command string. You may have 
inserted a character such as #, $, or % in the command header or within a 
parameter. Example: TRIG:SOUR BUS#
-102
Syntax error 
Invalid syntax was found in the command string. You may have inserted a 
blank space before or after a colon (or before a comma) in the command 
header. Example: APPL:SIN   ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used 
a comma instead of a colon, semicolon, or blank space – or you may have 
used a blank space instead of a comma. Example: TRIG:SOUR,BUS
or  APPL:SIN 1  1000
-105
GET
 not allowed
A Group Execute Trigger (GET) is not allowed within a command string.