Agilent Technologies E1441A Manuale Utente

Pagina di 206
156
Agilent E1441A Function Generator Tutorial  
Appendix C
Phase Truncation Harmonics
 ≤  –
20 x log
10
(P)
dBc 
where “P” is the number of waveform points in RAM.
Output Amplitude Control
The Agilent E1441A uses a 12-bit digital-to-analog converter (DAC) to convert the 
digital representation of a signal to an analog output voltage. The 
DAC
 can create 
waveforms represented by 4,096 (2
12
) discrete amplitude levels. All 4,096 amplitude 
codes are used for the built-in waveforms. Output levels from full maximum to 
minimum output are controlled by applying varying amounts of signal gain or 
attenuation to the signal created by the 
DAC
 as shown in the block diagram below. 
The output waveform is always described by the full 12-bit vertical resolution. You 
can download user-defined arbitrary waveforms using less than the full 12-bit 
vertical resolution; however, it is recommended that you always use the full 
DAC
 
amplitude resolution to minimize amplitude quantization errors as previously 
discussed.
The Agilent E1441A has a fixed output source resistance of 50 ohms (see next 
diagram).  During calibration, output amplitudes are calibrated for both the 
open-circuit voltage (no load) and the terminated output voltage (loaded). The 
terminated output amplitude is calibrated for an exact 50 ohm load. Since the 
function generator's output resistance and the load resistance form a voltage divider, 
the measured output voltage of the function generator will vary with load resistance 
value and accuracy. When the function generator's output is loaded with a 0.2% 
accuracy termination, an additional (negligible) 0.2% amplitude error is created. 
Using a 5% accuracy termination will add 5% additional error to specified output 
amplitudes.
If the function generator's output is measured with no load connected, the output will 
be approximately twice the displayed amplitude (V
gen
instead of V
load
). In some 
applications, you might continually use the function generator in a “no-load” 
conditions. In such applications, remembering to double the function generator's 
displayed amplitude can cause many errors. The Agilent E1441A allows you to 
specify the function generator's load condition using the OUTPUT:LOAD 
command; thus enabling the function generator to display the correct output 
amplitude.
Figure C-4. 
Figure C-5.