Use the following command to set the burst phase:
BM:PHASe <degrees>|MINimum|MAXimum
Frequency-Shift
Keying (FSK)
Modulation
You can configure the function generator to “shift” its output frequency 
between two preset values using 
FSK
 modulation. The rate at which the 
output shifts between the two frequencies (called the “carrier frequency” and 
the “hop frequency”) is determined by the internal rate generator or the 
signal level on the"Ext Trig/FSK/Burst terminal". The function generator 
can produce an FSK waveform using sine, square, triangle, ramp, and 
arbitrary waveforms. 
Topics covered on frequency-shift keying modulation 
are:
“FSK Carrier Frequency” on page 40
“FSK “Hop” Frequency” on page 41
“FSK Rate” on page 41
“FSK "Hop" Source” on page 42
-- “Internal Hop Source” on page 42
-- “External Hop Source” on page 42
Only one modulation mode can be enabled at a time. When you enable 
FSK
the previous modulation mode is turned off. 
Use the following command to enable frequency-shift keying: To  ensure 
proper operation, you should enable 
FSK
 after you have set up the other 
modulation parameters.
FSKey:STATe OFF|ON
FSK Carrier Frequency
FSK Carrier frequency:
10 mHz to 15 MHz  (100 kHz for triangle 
and ramp). The default is 1 kHz. You can use sine, square, ramp, triangle, or 
arbitrary waveforms for the carrier waveform.
For arbitrary waveforms, the maximum carrier frequency depends on the 
number of points specified in the waveform. The five built-in arbitrary 
waveforms can be output at a maximum of 5 MHz.    
For an externally-controlled 
FSK
 waveform, the carrier frequency is output 
when a "low" 
TTL
 level is applied to the source selected by the TRIG:SOUR 
command. The hop frequency is output when a “high” 
TTL
 level is applied.
For 
FSK
, the sync signal is referenced to the 
FSK
 “hop” signal (not the 
carrier). A momentary 
TTL
 “high” pulse (> 200 ns) is output on the transition 
to the “hop” frequency. The signal is output from the front- panel Sync 
terminal. 
Table 2-8. 
Number of Arb Points
Minimum Frequency
Maximum Frequency
8 to 8,192 (8k)
100 mHz
5 MHz
8,193 to 12,287 (12k)
100 mHz
2.5 MHz
12,288 to 16,000
100 mHz
200 kHz