Agilent E1441A SCPI Command Reference  
87
Chapter 3
OUTPut
 Subsystem
Syntax
OUTPut
:LOAD
50 | INFinity | MINimum | MAXimum
:LOAD?
MINimum | MAXimum
:SYNC
OFF | ON
:SYNC?
:TRIGger:IMMediate
:TRIGger:STATe OFF | ON
:TRIGger:STATe?
:TTLTrg<n>[:STATe]  <mode>
:TTLTrg<n>[:STATe]?
:LOAD
OUTPut:LOAD 50|INFinity|MIN|MAX 
selects the output termination for output 
amplitude and offset voltage. The function generator has a fixed output impedance 
of 50 ohms on the "Output" terminal. You can specify whether you are terminating 
the output into a 50 ohm load or an open circuit. Incorrect impedance matching 
between the function generator and your load will result in an amplitude or offset 
which does not match the specified signal level. The output load setting is stored in 
volatile memory.
Parameters
Comments
The amplitude (or dc offset) is automatically adjusted if you change the output 
termination (and no error is generated). For example, if you set the amplitude 
to 10 Vpp and then change the termination from 50 ohms to “high impedance”, 
the amplitude will double to 20 Vpp. If you change from “high impedance” to 
50 ohms, the amplitude will drop in half.
If you specify a 50 ohm termination but are actually terminating into an open 
circuit, the output will be twice the value specified. For example, if you set the 
offset to 100 mVdc (and specify a 50 ohm termination) but are actually 
terminating the output into an open circuit, the actual offset will be 200 mVdc.
*RST Condition:  OUTPut:LOAD 50
:LOAD?
OUTPut:LOAD? [MIN|MAX] 
queries the output impedance. Returns “50” or 
“9.9E+37”.
Parameter Name
Sets output termination to:
50 (default)
50
Ω
INFinity
"high impedance"
MIN
50
Ω
MAX
"high impedance"