Agilent Technologies E1441A Manuale Utente

Pagina di 206
90
Agilent E1441A SCPI Command Reference  
Chapter 3
PHASe 
Only valid with
Option 001
The PHASe command subsystem is valid only with the Phase-Lock option (option 
001).  This command allows you to control the phase offset of the output waveform. 
You can also enable the phase-lock error reporting to generate an error when 
phase-lock is lost (unlocked).  See the STATus command for querying the 
questionable data registers and status byte to verify the unlock condition.
Subsystem Syntax
 PHASe
:ADJust <radians> | <degrees> DEG | MINimum | MAXimum
:ADJust?
:REFerence
:UNLock:ERRor:STATe OFF | ON
:UNLock:ERRor:STATe?
:ADJust
PHASe:ADJust <radians>|<degrees>DEG |MIN|MAX 
adjusts the phase offset of 
the output waveform in radians or in degrees (must specify DEG following the 
parameter if specified in degrees).  Select from -2
π radians to +2π radians.  The 
default is 0 radians.  MIN = -2
π radians; MAX = +2π radians.  The phase offset 
setting is stored in volatile memory. 
Parameters
Comments
You must specify "DEG" following the phase value if you specify the phase in 
degrees instead of radians.  For example, to specify a -90
°  phase offset, use the 
following command:
PHAS:ADJ -90 DEG
For sine, square, triangle, and ramp waveforms:
Zero (0) radians is the point at which the waveform crosses zero volts (or the 
dc offset value) in a positive-going direction.
For arbitrary waveforms:
Zero (0) radians is the first point downloaded to memory.
The phase adjustment for phase-lock is independent of the burst phase as set by 
the 
BM:PHAS
 command.  
See “Burst Modulation” on page 33.
for more 
information.
Parameter Name
Parameter Type
Range of Values
Default Units
radians
numeric
-2
π
  to +2
π
 radians
degrees
numeric
-360
°
 to +360
°
must specify 
DEG