Agilent E1441A SCPI Command Reference  
Chapter 3
APPLy Commands
The APPLy command provides the most straightforward method to program the 
function generator.  It allows you to select the output function, output frequency, 
output amplitude and a dc offset voltage on which the output signal resides in a single 
command.
Note
The APPLy commands not only set the instrument’s function, frequency, 
amplitude, and offset, it also sets Duty Cycle to 50%, turns Modulation 
OFF, sets the TRIG:SOUR to IMM, and sets the TRIG:SLOPE to POS. 
Because of this, where a sequence of commands is needed to configure the 
instrument, and one of them will be a SOUR:APPLY command, the APPLy 
command must be first.
APPLy Command Parameters
Each APPLy command specifies the function to be output followed by three 
parameters. 
Subsystem
Syntax
APPLy:<FUNCTION> [<frequency> | DEFault[,<amplitude> | DEFault[,<offset>]]]
Syntax Example
Executing the following statement will output a sine wave with frequency of 5 kHz 
and amplitude of 3 Vpp with a -2.5 volt offset:
"APPL:SIN 5.0E+3, 3.0, -2.5"
default units are hertz, Vpp, VDC
Because of the use of optional parameters in the APPLy commands (enclosed in 
square brackets), you must specify frequency to use the amplitude parameter, and 
you must specify both frequency and amplitude to use the offset parameter. The 
following statement is valid (frequency and amplitude are specified; offset is 
omitted):
"APPL:SIN 5.0E+3, 3.0"
However, you 
CANNOT
 specify amplitude or amplitude and offset without specifying 
frequency.
Parameter
Name
Parameter
Type
Range of
Values
Default
Setting
frequency
numeric
100 
µ
Hz - 15 MHz
(see the following section 
titled "Output Frequency")
1 kHz
amplitude
numeric
50 mVpp - 20 Vpp
(see the following section 
titled "Output Amplitude")
100 mVpp
(into 50 ohms)
offset
numeric
see equation below
 0 Vdc