42
CFA3000 MANUAL
SECTION 7
INSTRUMENT DESCRIPTION
Port
Function
RS­232­C
For connection of 
computer interface or  
other accessories
Relay Output
Optional relay contact   
(multi­stream analyzer  
only)
Printer
For connection of printer  
cable
Recorder
For connection of
external recorder; for  
multi­output, 4­20 mA  
and 0­5 Vdc outputs for  
streams 2 through 6.
Pin No.
Description
1 NO    9 C
Stream #1 Alarm or ID
2 NO  10 C
Stream #2 Alarm or ID
3 NO  11 C
Stream #3 Alarm or ID
4 NO  12 C
Stream #4 Alarm or (#1*) ID
5 NO  13 C
Stream #5 Alarm or (#2*) ID
6 NO  14 C
Stream #6 Alarm or (#3*) ID
7.6
Interface Ports
The interface ports (labeled for recorder, printer,
relay output, and RS­232) are located on the lower,
left outside wall.
Table 7­12:    Interface Ports
If there are 3 streams or less, both stream ID and
alarms may be optionally configured.
7.8 Multi­Output Board
The output for Stream #1 is obtained from the
Input/Output Panel (see Figure 2­5). The additional
five outputs are obtained from the lower left­hand side
of the analyzer. Table 7­14 gives the pin assignments
on the “D” connector for a multi­stream system: 
Figure 7­6.  Interface Ports
7.7     Relay Output Option
This port (see Figure 7­7) is a collection of contact
closures that can be used to identify either the
stream being displayed or the alarm that is activated
(multi­stream systems only). these are not “powered”
contacts.
Relay Contact Ratings:
100 Vdc ­­ 
max voltage
10   Vdc­­
max DC current
1.0  A ­­ 
max carry current
0.5  A ­­ max switched current
Table 7­13:  Relay Output Option
Pin Assignments
1  2  3  4  5  6  7  8  9  10  11  12  13  14 /
\15  16  17  18  19  20  21  22  23  24  25 /
Figure 7­7. "D" Connector for Relay Output
Table 7­14:  Multi­output Pin Assignments on 
“D” Connector  
2
15
2
14
1
3
17
4
16
3
4
19
6
18
5
5
21
8
20
7
6
23
10
22
9